OpSens’ OTG-M Series fiber optic temperature sensors (OTG-M170, OTG-M220, OTG-M360, OTG-M420, OTG-M600, OTG-MPK5, and OTG-MPK8) provide outstanding performance. The OTG-M Series temperature sensors use a tried-and-tested temperature transduction framework based on the temperature-dependent bandgap of Gallium Arsinide (GaAs) crystal.
Due to the tiny GaAs crystal situated at the tip of an optical fiber, this sensor type is useful for tip measurement systems. Each model of the OTG-M series is distinguished by its response time and sensor diameter (OD).
When merged with OpSens’ SCBG signal conditioning technology and the advantages of measuring with light, the OTG-M Series provides unprecedented continuity and consistency in the harshest conditions, like elevated levels of EM, RF, MR, and microwave field settings.
The OTG-M Series is built with industry-standard optical fiber and is compatible with all OpSens SCBG signal conditioners. This compact and dependable fiber optic temperature sensor can be customized for OEM or customer-specific systems, and it is available with a range of optical cables and sheath options.
Key features
- Good accuracy
- Intrinsically secure
- OEM and tailored versions
- Resistant to EMI, RFI, MRI, and microwave radiation
- Strong and compact design with excellent repeatability
Applications
- High-voltage settings
- Microwave, RF, and EM settings
- Applications that need compact sensors
- Observing temperature variations
- Disposable sensor, low-cost
OTG-M series
Fiber optic temperature sensor
Image Credit: OpSens Medical
Specifications
Source: OpSens Medical
|
Sensor model |
OTG-M170 |
OTG-M220 |
OTG-M360 |
OTG-M420 |
OTG-M600 |
OTG-MPK5 |
OTG-MPK8 |
Response time |
≤35 ms |
≤50 ms |
<100 ms |
<200 ms |
200 ms |
<225 ms |
<500 ms |
Sensor diameter |
0.170
mm |
0.220
mm |
0.360
mm |
0.420
mm |
0.600
mm |
0.510
mm |
0.810
mm |
Sensor tip material (Tube) |
Bare fiber |
Polyimide |
Glass |
Polyimide |
Polyimide |
PEEK |
PEEK |
Temperature operating and calibrated range |
-40 °C to +250 °C (other ranges available upon request) |
Accuracy |
±0.8 °C or better for other temperature range |
Operating humidity range |
0-100% |
EMI/RFI Susceptibility |
Complete immunity |
Calibration |
NIST traceable |
Cable length |
1.5 meters standard (other lengths available) |
Optical connector |
LCA, ST, standard |
Cable sheathing |
TeflonTM |
Signal conditioner compatibility |