The multipurpose, high-speed Avanti® J-26 XP, capable of processing up to six liters in less than ten minutes, is an ideal solution for multi-user, multi-protocol labs.  

With a full range of rotors, adapters and sample carriers, the Avanti-26 XP Series offers incredible versatility to support your research needs for many years to come.

Features

A single J-26 XP can process six liters of fermentor output in less than ten minutes, deliver a subcellular fraction at 82,000 x g, run 24 microplates at a time and separate live cell populations with elutriation.

  • Shorter cycle times and no waiting for your sample at the end of a run  with high-torque, switched reluctance (SR)* technology
  • User-defined acceleration and deceleration rates for improved efficiency, yield and confidence
  • Full containment barrier, automatic rotor identification and enhanced biosafety for greater lab safety
  • Extensive rotor library including lightweight carbon fiber designs, fixed-angle, swinging bucket and elutriation**
  • Compatible with J-Lite rotors
  • Quieter, cooler and less energy use than traditional centrifuges

Avanti J-26 XP Centrifuge System Specifications

UNSPSC

41103905

Height

86 cm (34”); with door open 149 cm (58.7”)

Width

71 cm (28“)

Depth

86 cm (34”); 102 cm (40.25”) (includes air diverter extending from back panel)

Weight

290 kg (640 lbs)

Friction Reduction System (FRS)

Mechanical rotary vane pump reduces chamber pressure to 0.25 ATM (190 mm [7.5”] Hg)

Max capacity volume range

6 liters

Max speed

26,000 rpm

Noise level (0.91meters /3' in front of centrifuge)

< 57 dbA

Max heat dissipation into room (steady-state conditions)

2.0 kW (6,900 Btu/hour)

Rotor ID

Manual

Set temperature

-10° C to +40° C (in 1° C increments)

Set time

1 min-3 hours, Hold

Refrigeration system

Non-CFC, non-ozone depleting refrigerant

Speed control

± rpm of set speed or 0.1%, whichever is greater

Speed range

100 to 26,000 rpm or equivalent RCF; 0 to 5,000 rpm for elutriation

Temperature control

± 2° C of set temperature (after equilibration)

Zonal capability

No

Single tube sample volume range

1.5 - 1,000 mL

Ambient temperature range

15° C to 40° C

Clearances required

7.6 cm (3.0”) sides; 16.0 cm (6.25”) rear

Door

6.1 cm (2.4”) thick structural foam with steel plate

Humidity restrictions

< 95% (non-condensing)

Pollution degree

2

Rotor chamber diameter

51.3 cm (20”)

Time–actual display

Indicates run time remaining (timed run) or elapsed time (hold run)

Maximum force at r max (g)

82,000 x g

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.