The multipurpose, high-speed Avanti® J-26 XP, capable of processing up to six liters in less than ten minutes, is an ideal solution for multi-user, multi-protocol labs.
With a full range of rotors, adapters and sample carriers, the Avanti-26 XP Series offers incredible versatility to support your research needs for many years to come.
Features
A single J-26 XP can process six liters of fermentor output in less than ten minutes, deliver a subcellular fraction at 82,000 x g, run 24 microplates at a time and separate live cell populations with elutriation.
- Shorter cycle times and no waiting for your sample at the end of a run with high-torque, switched reluctance (SR)* technology
- User-defined acceleration and deceleration rates for improved efficiency, yield and confidence
- Full containment barrier, automatic rotor identification and enhanced biosafety for greater lab safety
- Extensive rotor library including lightweight carbon fiber designs, fixed-angle, swinging bucket and elutriation**
- Compatible with J-Lite rotors
- Quieter, cooler and less energy use than traditional centrifuges
Avanti J-26 XP Centrifuge System Specifications
UNSPSC
|
41103905
|
Height
|
86 cm (34”); with door open 149 cm (58.7”)
|
Width
|
71 cm (28“)
|
Depth
|
86 cm (34”); 102 cm (40.25”) (includes air diverter extending from back panel)
|
Weight
|
290 kg (640 lbs)
|
Friction Reduction System (FRS)
|
Mechanical rotary vane pump reduces chamber pressure to 0.25 ATM (190 mm [7.5”] Hg)
|
Max capacity volume range
|
6 liters
|
Max speed
|
26,000 rpm
|
Noise level (0.91meters /3' in front of centrifuge)
|
< 57 dbA
|
Max heat dissipation into room (steady-state conditions)
|
2.0 kW (6,900 Btu/hour)
|
Rotor ID
|
Manual
|
Set temperature
|
-10° C to +40° C (in 1° C increments)
|
Set time
|
1 min-3 hours, Hold
|
Refrigeration system
|
Non-CFC, non-ozone depleting refrigerant
|
Speed control
|
± rpm of set speed or 0.1%, whichever is greater
|
Speed range
|
100 to 26,000 rpm or equivalent RCF; 0 to 5,000 rpm for elutriation
|
Temperature control
|
± 2° C of set temperature (after equilibration)
|
Zonal capability
|
No
|
Single tube sample volume range
|
1.5 - 1,000 mL
|
Ambient temperature range
|
15° C to 40° C
|
Clearances required
|
7.6 cm (3.0”) sides; 16.0 cm (6.25”) rear
|
Door
|
6.1 cm (2.4”) thick structural foam with steel plate
|
Humidity restrictions
|
< 95% (non-condensing)
|
Pollution degree
|
2
|
Rotor chamber diameter
|
51.3 cm (20”)
|
Time–actual display
|
Indicates run time remaining (timed run) or elapsed time (hold run)
|
Maximum force at r max (g)
|
82,000 x g
|