Medtronic informs physicians about the incorrect labeling of current SC catheter

NewsGuard 100/100 Score

Medtronic, Inc. (NYSE: MDT) today announced that the U.S. Food and Drug Administration (FDA) has classified its letters to physicians related to the incompatibility of its Sutureless Connector (SC) intrathecal catheters with the IsoMed® constant-flow infusion pumps as a Class I recall. In the letters, Medtronic informed physicians that the current SC catheter labeling incorrectly states that SC catheters are compatible with the IsoMed infusion pump.

The affected products include the following models:
SC Catheter and Revision Kit Models: 8709SC, 8731SC, 8578, and 8596SC when paired with the IsoMed Pump Model 8472, which is no longer manufactured and marketed. Medtronic stopped producing IsoMed infusion pumps in September 2008 as part of a planned product phase-out. Other types of intrathecal catheters and intravascular catheters for delivery of chemotherapy are not affected by this issue.

A physical interference between the SC catheter connector and the IsoMed pump prevents the SC catheter from completely connecting to the IsoMed pump, even though it may appear to be connected and feel secure. This incompatibility is design related and is not a result of implant technique. This incompatibility may lead to disruptions of therapy and revision surgery, which pose a risk of serious injury or death.

The clinical manifestations of an incomplete connection of an SC catheter to an IsoMed pump may include, but are not limited to, drug or cerebrospinal fluid (CSF) leakage into surrounding tissue, resulting in tissue damage, loss of or change in therapy, a return of underlying symptoms, drug withdrawal symptoms, a severe (“spinal”) headache, and/or a clinically significant or fatal drug underdose. Patients receiving intrathecal baclofen therapy are at higher risk for adverse events as baclofen withdrawal can lead to a life threatening condition if not treated promptly and effectively. The IsoMed pump is not FDA approved in the United States for the administration of baclofen.

To date, Medtronic has received ten reports worldwide related to improper connection of an SC catheter to an IsoMed pump. In all ten reports, medical intervention was required to correct the condition. The ten reported events included reports of underlying symptoms, withdrawal symptoms and fluid collection in the pump pocket. In one case, it was reported that a patient death occurred two days following device revision, but it cannot be determined whether the cause of death is or is not related to the device.

The recommendations provided to physicians included directions not to use the SC catheters with an IsoMed pump. If an SC catheter is, or has been, connected to an IsoMed pump:

  • Clinical judgment should be used to determine if prophylactic revision is warranted based on a patient’s risk for a return of symptoms or complications due to drug withdrawal and the surgical risk of catheter revision.
  • The SC catheter must not be reconnected to any other pump because the SC catheter connector may have been permanently damaged.

Medtronic is in the process of communicating with physicians and patients about this issue. Medtronic reminds patients to immediately seek medical attention if they experience a return of underlying symptoms or drug withdrawal symptoms.

http://www.medtronic.com/

Comments

The opinions expressed here are the views of the writer and do not necessarily reflect the views and opinions of News Medical.
Post a new comment
Post

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.

You might also like...
AI technologies can accurately identify cases of healthcare-associated infections