New 'leadless' pacemakers less likely to cause complications, research finds

NewsGuard 100/100 Score

Research finds use of smaller, self-contained devices reduced complications for up to 18 months compared with conventional designs

Pacemakers that don’t use wires to connect the device to the heart are successfully reducing the number of short-term and mid-term complications patients experience, a study in the United States has found.

The findings of the research, led by Ohio-based Cleveland Clinic, appeared in the journal Heart Rhythm. The study compared patient outcomes for those fitted with new ‘leadless’ pacemakers compared with more conventional ‘transvenous’ designs.

Lead author Dr. Daniel Cantillon, M.D., research director for Cardiac Electrophysiology and Pacing at Cleveland Clinic, and a consultant for Abbott and Boston Scientific, said:

The data from this study is encouraging, and we expect complications from leadless pacemakers to continue to decline as the technology improves and physicians gain experience implanting these devices.

While this research shows benefit for leadless pacing, we must keep in mind that the field is still too young to compare the long-term results of this technology, the implications of which will not be fully understood for at least another decade."

More than one million pacemakers are implanted annually worldwide, providing electrical stimulation to regulate a patient’s heartbeat. Conventional models are surgically placed under the skin of the chest, and connected to the heart via a wire that stretches from the shoulder vein to the heart. These leads – and their surgical implantation – are the most common source of complications among pacemaker recipients as a whole, occurring in up to 12 percent of this group, according to previous research.

Leadless pacemakers, by contrast, do not need wires. These small, self-contained devices – about one-tenth of the size of a traditional pacemaker – are placed directly into the heart using a catheter that is carried from the leg to the heart via the thigh’s femoral vein. The first leadless pacemaker was introduced in 2014 and approved by the U.S. Food & Drug Administration (FDA) two years later.

The new multicenter study compared short- and mid-term complications between 718 patients receiving the Nanostim leadless pacemaker and 1,436 patients with conventional (transvenous) pacemakers. Leadless pacemaker patient data was taken from the LEADLESS II trial, a prospective, nonrandomized, multicenter clinical trial. Transvenous patient data were obtained from Truven Health MarketScan claims databases for patients implanted with single-chamber pacemakers between April 2010 and March 2014 and more than one year of pre-implant enrollment data. Statistical methods were used to match patients between the two groups to compare the outcomes of a leadless versus traditional pacemaker with other key clinical variables being equal.

At one month, the study found that patients receiving one type of leadless pacemaker (Nanostim) overall had fewer complications (5.8 percent versus 9.4 percent). Leadless pacemakers eliminated lead and pocket complications, including infection. By comparison, complications among traditional pacemaker recipients included lead complications (3.62 percent), pocket complications (0.42 percent) and infection (1.74 percent).

There were no significant differences between the groups in regard to rates of vascular complications, electrode dislodgement, or generator complications. However, the study did find that those receiving leadless pacemakers had an increased risk of developing pericardial effusion – bleeding between the heart and the sac that surrounds it – of 1.53 percent versus 0.35 percent. These complications were uncommon but serious, and sometimes required surgery.

Beyond one month and up to 18 months of follow-up, leadless patients continued to experience overall fewer complications than transvenous patients (0.56 percent versus 4.94 percent). In the conventional pacemaker group, a number of complications were wholly absent from the leadless group, including lead-related complications, electrode dislodgement, infection, and pocket complications.

Comments

The opinions expressed here are the views of the writer and do not necessarily reflect the views and opinions of News Medical.
Post a new comment
Post

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.

You might also like...
Research confirms no association between SARS-CoV-2 and childhood asthma diagnoses