Electron-beam sterilized dialysis membrane may increase risk of thrombocytopenia

NewsGuard 100/100 Score

Patients who had undergone hemodialysis using dialyzers that had been sterilized with the use of electron beams were more likely to develop thrombocytopenia (an abnormally low platelet count in the blood, associated with increased risk of bleeding), according to a study in the October 19 issue of JAMA.

"Adverse device reactions to hemodialysis treatments are uncommon but can still occur in today's era of hemodialysis membranes and technology. During hemodialysis treatment, patients are exposed to a variety of components of the dialysis circuit and could have an adverse reaction to blood tubing, dialyzer membranes [used in the dialysis machines], and dialysis solutions," according to background information in the article. "Thrombocytopenia is not widely recognized as a potential dialyzer-related complication. Following the observation of significant thrombocytopenia among 20 patients undergoing hemodialysis in a single dialysis unit after the introduction of dialyzers sterilized by electron beam (e-beam), a larger investigation was undertaken."

Mercedeh Kiaii, M.D., F.R.C.P.C., of the University of British Columbia, Vancouver, Canada, and colleagues conducted a study to determine the prevalence and cause of thrombocytopenia in patients undergoing hemodialysis in British Columbia and southern Alberta province between April 2009 and November 2010. Retrospective analyses of historical patient, laboratory, and dialyzer data predating conversion to e-beam dialyzers were undertaken, with prospective collection of predialysis and postdialysis platelet counts before and after the change from e-beam to non-e-beam sterilized dialyzers in September 2009. Significant thrombocytopenia was defined as post-dialysis treatment platelet count of less than 100 x 103/µL and a postdialysis decrease in platelet count of more than 15 percent.

Province-wide testing of all patients from British Columbia who underwent hemodialysis in community and hospital-based facilities permitted collection of data from the majority (80 percent) of patients with hemodialysis from British Columbia (1,706 of 2,146 patients) and from Alberta (425 patients). Of the 1,706 patients from British Columbia, 1,411 patients (83 percent) were undergoing dialysis using e-beam sterilized membranes. In British Columbia, 11.4 percent of patients (n = 194) had postdialysis platelet counts of less than 100 x 103/µL, 23.4 percent of patients (n = 400) had more than 15 percent decrease in platelet count postdialysis, and 7.2 percent of patients (n= 123) met both criteria. Similar proportions from the Alberta population who were undergoing dialysis using a polysulfone e-beam sterilized dialyzer made by a different manufacturer were demonstrated (10.8 percent; n=46 had postdialysis platelet counts of less than 100 x 103/µL; 32.0 percent; n= 156 had more than 15 percent decrease in platelet count postdialysis; and 7.3 percent (n = 31) met both criteria.

In multivariate analysis adjusting for patient and dialysis history characteristics, use of an e-beam sterilized dialyzer was associated with a 2.5 times increased risk of significant thrombocytopenia. "Overall, among 1,784 patients, there were significant reductions in postdialysis thrombocytopenia following the change to use of non-e-beam sterilized dialyzers, such that 120 patients (6.7 percent) had platelet counts of less than 100 x 103/µL; 167 patients (9.4 percent) had decreases in platelet counts of greater than 15 percent; and 38 patients (2.1 percent) met both criteria," the authors write.

"Our study highlights 2 important points in the care of patients undergoing hemodialysis. First, dialyzer reactions such as thrombocytopenia can occur even with the use of current technologically advanced dialyzer membranes and devices. Second, these reactions may be overlooked with current routine predialysis blood work. Hemodialysis unit protocols and guidelines might consider adding postdialysis blood counts to routinely performed blood work when new dialyzers are introduced to help identify possible dialyzer-associated adverse effects," the authors write.

The researchers add that it is biologically plausible that e-beam sterilization could induce changes in platelets by a number of possible mechanisms, but none of these have been thoroughly evaluated.

In an accompanying editorial, Jonathan Himmelfarb, M.D., of the University of Washington, Seattle, comments on the findings of this study.

"This systematic, multifaceted study clearly demonstrated a robust association of thrombocytopenia risk with use of e-beam sterilization of dialyzers. It is also an exemplary model for investigating other problems that can affect patient safety during complex procedures such as dialysis. Several recently published single-patient case reports also have implicated e-beam sterilization as a possible cause of dialysis-induced thrombocytopenia."

Source: JAMA

Comments

The opinions expressed here are the views of the writer and do not necessarily reflect the views and opinions of News Medical.
Post a new comment
Post

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.

You might also like...
Genetic analysis reveals hidden causes of chronic kidney disease in adults