Positive data from Echo's Symphony tCGM System clinical trial on type 1 or 2 diabetes

NewsGuard 100/100 Score

Echo Therapeutics, Inc. (Nasdaq: ECTE), a company developing its needle-free Symphony® tCGM System as a non-invasive, wireless, transdermal continuous glucose monitoring system and its Prelude® SkinPrep System for transdermal drug delivery, presented an expanded analysis of data collected from subjects with either type 1 or type 2 diabetes during its recent clinical trial of its Symphony tCGM System.  The data were presented during an oral presentation session on Friday, June 8 at the 72nd Annual Scientific Sessions of the American Diabetes Association (ADA) in Philadelphia, PA.

The presentation [ADA abstract 007-OR], entitled "Transdermal Continuous Glucose Monitoring Following Skin Permeation," by Wayne Menzie, Echo's Director of Technology and Clinical Development, provided data from twenty (20) adult subjects with type 1 or type 2 diabetes whose glucose levels were monitored continuously for 24 hours.  The key findings include:

  • Symphony successfully monitors the broad range of blood glucose values seen in people with diabetes.
  • Using over 2,600 Symphony tCGM glucose readings paired with reference blood glucose measurements, the Mean Absolute Relative Difference (MARD) was 12.6% and the Median Absolute Relative Difference (MedianARD) was 11.8%.  The overwhelming majority of sensors had a MARD between 8.0% and 15.9% with an overall MARD of 11.3% and a MedianARD of 10.7%.
  • The Continuous Glucose-Error Grid analysis (CG-EGA) showed that 94.4% of the readings were clinically accurate and 2.5% were benign errors with a combined A+B of 96.9%.
  • Values for blood glucose measurements ranged from 42 to 333 mg/dL.  Most of the study subjects experienced a large degree of glycemic variability with twelve (12) having a standard deviation of all blood glucose measurements of greater than 40 mg/dL and only four (4) subjects less than 30 mg/dL.
  • An important component of continuous glucose meter performance, the R deviation is a measurement of rate accuracy that examines the difference between the rate of the reference blood glucose measurements and the calculated rate of the CGM over that same time.  In this study, 84.7% of the calculated CGM rates were within + or - 1 mg/dL/min of the reference rate and 96.8% were within + or - 2 mg/dL/min. 

Following the oral presentation, Echo hosted a panel discussion on hospital-based glucose control with Key Opinion Leaders (KOLs) in glycemic control.   Anthony Furnary, M.D. (Starr-Wood Cardiac Group, Portland, OR), Jeffrey Joseph, D.O. (Thomas Jefferson University Hospital, Philadelphia, PA), Stanley Nasraway, M.D. (Tufts Surgical Center, Boston, MA), and Eileen Donnelly, R.N., B.S.N. (Thomas Jefferson University Hospital, Philadelphia, PA) discussed the need for continuous glucose monitoring in hospitals.

"We are extremely pleased with the outcome of the study and our Research & Development team's progress on Symphony.  We believe that Symphony successfully monitors glucose in the challenging diabetes population who experience rapid rates of change and a broad range of glucose levels," said Patrick T. Mooney, M.D., Chairman and Chief Executive Officer of Echo Therapeutics. "Additionally, the KOLs at our symposium confirmed the major need for hospital-based continuous glucose monitoring to better control glucose levels in critically ill patients.  We value this opportunity for engagement with the medical community to share the Symphony diabetes and critical care data with experts in the field."

Source:

Echo Therapeutics, Inc.

Comments

The opinions expressed here are the views of the writer and do not necessarily reflect the views and opinions of News Medical.
Post a new comment
Post

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.

You might also like...
Advanced insulin pumps enhance blood sugar management in type 1 diabetes