DRE Waveline Pro Touch-Screen Anesthesia Monitor

Description

Superior Performance with a High-Resolution Touch-Screen

The DRE Waveline Pro displays as many as eight colorful waveforms on a crisp, high-resolution 15-inch touch-screen that optimizes the speed of patient care. The user can make quick screen adjustments, set default settings, alarm limits, and manage up to 72 hours of detailed patient data.

Designed for a fast paced work environment, the Waveline Pro patient monitor offers an extremely simple and adaptable user interface. Patient information along with vital-signs settings can be quickly modified to meet the needs of a patient's changing condition, converting from a basic vital signs monitor, to a continuous bed side monitor, to an operating room monitor. Now with Masimo SET® Pulse Oximetry!

This patient monitor offers standard measurements of: non-invasive blood pressure, 3-lead or 5-lead ECG with arrhythmia detection, motion tolerant SpO2, Temperature, and Respiration rate. End-tidal CO2, Anesthetic Agent measurement, Cardiac Output and Invasive blood pressure are available options.

The Waveline Pro monitor offers several connective solutions to network multiple monitors and/or manage patient data on an electronic medical records platform or a HL7-based hospital information system. It offers Ethernet and RS-232 connections with an open source communication protocol, and it is HL7 compliant. The HL7 network protocol will allow for all patient information and vital sign trends to be transferred and stored on a hospital information system.

Capnography is outpacing Pulse Oximetry as the new standard of care in outpatient surgery procedures.

Features

  • New - Masimo SET® Pulse Oximetry — ask your DRE representative for details
  • Optional Integrated Dräger Anesthesia Gas Bench expands measurement capabilities
  • Optional agent bench is compact and part of main unit. All in one monitor!
  • Large (15"), touch-screen with high-resolution flat panel display
  • Displays as many as eight waveforms
  • Multi-lead simultaneous ECG monitoring
  • Automatically set alarm limits
  • Graphical and tabular trending
  • Color-coded alarms
  • Patient data entry
  • Networking capability
  • Works with EMR and wireless telemetry systems
  • Large time stamp in upper right corner
  • Standby button
  • Volume and Sounds are adjustable and can be turned off completely
  • Option to print Numerical Data Only
  • Color of the waveforms can be changed
  • Battery backup (two removable batteries)
  • Available with a built-in thermal printer

Specifications

Performance

  • Display: 15-inch color touch-screen
  • Trace: 8 waveforms
  • Indicator:
    • Alarm indicator
    • Power indicator
    • QRS beep and alarm sound
  • Trend time: 1 - 72 hour
  • Recorder:
    • Built-in, thermal array, 3 channels
    • Record width: 48 mm
    • Recorder paper: 50 mm
    • Record speed: 25 mm/s, 50 mm/s

ECG

  • Input: 5-lead ECG cable and standard AAMI line for connection
  • Lead Choice: I, II, III, aVR, aVF, aVL, V, V1-V6, TEST
  • Gain Choice : x0.5, x1, x2, x4
  • Frequency Characteristic: 0.05 ~ 35 HZ (+3 dB)
  • ECG Waveforms: 7 channels
  • Penetration Voltage: 4000 VAC 50/60 Hz
  • Sweep Speed: 12.5, 25, 50 and 100 mm/sec (left to right or right to left)
  • HR Display Range: 30 ~ 300 bpm
  • Accuracy: ±1 bpm or ±1%, whichever is greater
  • Alarm Limit Range Setting:
    • Upper limit 100 ~ 200 bpm,
    • Uower limit 30 ~ 100 bpm

RESP

  • Measure Method: RA-LL impedance
  • Range: 0 ~ 120 rpm
  • Accuracy: ±3 rpm
  • Alarm Limit Setting:
    • Upper limit 6 ~ 120 rpm,
    • Lower limit 3 ~ 120 rpm
  • Sweep Speed: 12.5, 25, 50 and 100 mm/sec (left to right or right to left)

NIBP

  • Measuring Technology: automatic oscillating measurement
  • Cuff Inflating: <30 s (0 ~ 300 mmHg, standard adult cuff)
  • Measuring Period: AVE<40s
  • Mode: Manual, Auto
  • Measuring Interval in AUTO Mode: 2 min ~ 4 hrs
  • Pulse Rate Range: 30 ~ 250 (bpm)
  • Measuring Range:
  • Adult/Pediatric Mode: SYS: 40 ~ 250 (mmHg)
  • DIA :15 ~ 200 (mmHg)
  • Neonatal Mode: SYS: 40 ~ 135 (mmHg)
  • DIA : 15 ~ 100 (mmHg)
  • Accuracy:
    • Maximum Mean error: ±5 mmHg
    • Maximum Standard deviation: 8 mmHg

TEMP

  • Range: 25 ~ 50 (°C)
  • Accuracy:
    • ± 0.2 °C (25.0 ~ 34.9 °C)
    • ± 0.1 °C (35.0 ~ 39.9 °C)
    • ± 0.2 °C (40.0 ~ 44.9 °C)
    • ± 0.3 °C (45.0 ~ 50.0 °C)
  • Display Resolution: 0.1 °C
  • Alarm Limit Setting:
    • upper limit 0 ~ 50 °C,
    • lower limit 0 ~ 50 °C
  • Channel: 2 channels

Masimo SET SpO2

  • SpO2 Accuracy (non-motion):
  • Adult Pediatric: 70~100%: ±2%, 0~69%: unspecified
  • Neonate: 70~100%: ± 3%,0~69%: unspecified
  • SpO2 Accuracy (motion):
  • Adult Pediatric: 70~100%: ±3%, 0~69%: unspecified
  • Neonate: 70~100%: ± 3%, 0~69% : unspecified
  • Low Perfusion Performance: >0.02% Pulse Amplitude and % Transmission >5
  • SpO2: ±2%
  • PR: ±3 bpm
  • Modes:
  • Averaging mode: 2,4,8,10,12,14 and 16 s
  • Sensitivity: Normal, APOD and Maximum
  • PR Accuracy (non-motion):
  • Neonate: 70~100%: ± 3%, 0~69  : unspecified
  • Adult Pediatric Neonate: 25~240 bpm: ±3 bpm
  • PR Accuracy (motion):
  • Adult Pediatric Neonate: 25~240 bpm: ±5 bpm
  • Measuring Range:
  • SpO2: 1~100%
  • PR: 25~240 bpm
  • Perfusion: 0.02~20%

IBP

  • Measurement Range: -50 ~ 300 mmHg
  • Channel: 2 channels
  • Pressure Transducer: sensitivity, 5 µV/V/mmHg
  • Impedance Range: 300 ~ 3000 ω
  • Transducer Sites: ART, PA,CVP, RAP, LAP, ICP
  • Unit: mmHg/kPa selectable
  • Resolution: 1 mmHg
  • Accurancy: ±1 mmHg or ±2%,whichever is greater
  • AlarmRange: -10 ~ 300 mmHg

EtCO2 (Optional)

  • CO2 Measurement Range: 0 ~ 99 mmHg
  • Accuracy:
    • ±2 mmHg (0 ~ 38 mmHg)
    • 39-99 mmHg ±5% of reading +0.08% for every 1mmHg (above 38 mmHg)
  • Sampling Rate: 50 ml/min
  • Initialization Time: 30 seconds (typical), reaches ±5% steady-state accuracy within 3 minutes.
  • Respiration Rate: 0 ~ 150 breaths/min
  • Mode: adult, neonate

Cardiac Output (Optional)

  • Measurement Method: Thermodilution Method
  • CO2 Measurement Range:
    • C.O.: 0.1 to 20 L/min
    • TB: 23 to 43
    • TI: 0 to 27
  • Resolution:
    • C.O.: ±5% or ±0.1 L/min, whichever is greater, as measured using electornically generated flow curves.
    • TB, TI: 0.1
  • Accuracy:
    • ±5 or ±0.1 L/min, whichever is greater, as measured using electronically generated flow curves.
    • TB, TI ±0.1 (without sensor)
  • Alarm Range: TB 23 to 43
  • Repeatability: C.O. ±2% or 0.1 L/min, whichever is greater, as measured using electronically generated flow curves.

Anesthetic Agents (Optional)

  • Method: Infrared absorption
  • Gas Sorts: Halothane, Isoflurane, Enflurane, Sevoflurane, Desflurane, CO2, N2O, O2 (optional Automatic Agent ID)
  • Measurement Range:
    • Halothane, Isoflurane: 0 - 8.5%
    • Enflurane, Sevoflurane: 0 - 10%
    • Desflurane: 0 - 20%
    • CO2: 0 - 10%
    • N2O: 0 - 100%
    • O2: 0 - 100%
  • Bias:
    • Halothane, Isoflurane, Enflurane, Sevoflurane, Desflurane: ±(0.15 Vol% + 15% rel.)
    • CO2: ±(0.5 Vol% + 12% rel.)
    • N2O: ± (2 Vol% + 8% rel.)
    • O2: ±3 Vol%

Networking

  • Industry standard 802.11 b/g wireless network

Power

  • Source: External AC power or internal battery
  • AC Power: 100 - 240 VAC, 50/60 Hz, 150 VA
  • Battery: Built-in and rechargeable lithium ion
  • Operating TIme: 3+ hours

Environmental

  • Temperature:
    • Operating: 5° - 40 °C
    • Storage: -20° - 65 °C
  • Humidity Range:
    • Operating: ≤;80%
    • Storage: ≤80%

Other Standard Feature

OxyCRG, drug dose calculation, cascading ECG, On screen NIPB trends (up to 250 readings), user set defaults, Arrhythmia detection, ST segment

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.