Using flow cytometry to support antibody discovery workflows

This article and associated images are based on a poster originally authored by Zaynab Isseljee, Michael Marshall, Sophie Snow, Adeola Onumonu, Claire Reader, Katy Everett, and Martina Trokter and presented at ELRIG Drug Discovery 2026 in affiliation with LifeArc.

This poster is being hosted on this website in its raw form, without modifications. It has not undergone peer review but has been reviewed to meet AZoNetwork's editorial quality standards. The information contained is for informational purposes only and should not be considered validated by independent peer assessment.

LifeArc's B-SMART™ antibody discovery platform

LifeArc is a non-profit medical research charity that supports early-stage research through partnerships, scientific expertise, and funding and investment to accelerate promising research for people living with rare diseases and drug-resistant infections.

The B-SMART™ antibody discovery platform will generate fully human antibodies for almost any viable target, ready for pre-clinical testing.

Flow cytometry is crucial to LifeArc's antibody discovery, enabling robust profiling of antibodies using cell-based assays and ensuring the delivery of antibodies with desired characteristics for further development.

Using flow cytometry to support antibody discovery workflows

Image Credit: Image courtesy of Zaynab Isseljee et al., in partnership with ELRIG (UK) Ltd.

1. Flow QC cell line target expression assay

Cell lines are engineered to stably express human, mouse, rhesus, or cynomolgus monkey (cyno) target(s) of interest during the feasibility phase.

  • Plasmids carrying the target of interest are transfected into relevant cells, the antibiotic is selected, expression is confirmed, and the cell lines are cryopreserved.
  • Flow cytometry is then used to confirm and validate each cell line's target expression level, including across passages (flow QC assay, Figure 1). Data enables the selection of cell lines for use in high-throughput (HT) flow cytometry-based screening assays.

CHO cells stably expressing human or cyno target of interest were tested in the Flow QC Cell Line Target Expression assay, using a positive control antibody that binds to the target and were detected with a secondary PE labelled antibody. Flow cytometry analysis was performed on the Attune NxT flow cytometer. Initial QC shows positive target expression in all cell lines generated compared to CHO parental and unstained cells. Stable expression levels are maintained in the majority of cell lines after 6 additional cell passages.

Figure 1. CHO cells stably expressing human or cyno target of interest were tested in the flow QC cell line target expression assay, using a positive control antibody that binds to the target and was detected with a secondary PE-labeled antibody. Flow cytometry analysis was performed on the Attune™ NxT flow cytometer.

Initial QC shows positive target expression in all cell lines generated compared to CHO parental and unstained cells. Stable expression levels are maintained in the majority of cell lines after six additional cell passages. Image Credit: Image courtesy of Zaynab Isseljee et al., in partnership with ELRIG (UK) Ltd.

2. Receptor quantification assay

Stable target-expressing cell lines are fully validated and characterized for use in the HT flow cytometry screening assays.

  • The receptor quantification assay is used to determine the absolute number of cell surface receptors per cell using flow cytometry (Figure 2).
  • Combining data from the flow QC cell line target expression assay and the receptor quantification assay strengthens decision-making in the selection of cell lines to take forward for use in the HT flow screening assays.

Source: LifeArc 

Cell Line MFI ABC
Human target cell line one 32,516 494,234
Human target cell line two 9345 165,775
Cyno target cell line one 5371 102,047
Cyno target cell line two 1213 27,713
Empty vector cells 46 N/A

CHO cells stably expressing human or cyno target of interest were tested in the Receptor Quantification Assay using the Quantum Simply Cellular® kit (Bangs Laboratories). Flow cytometry analysis was performed on the Attune™ NxT flow cytometer. The receptor quantification beads measure Antibody Binding Capacity (ABC) and were stained alongside cells with a positive control antibody; a standard curve is generated that enables interpolation of fluorescence intensity into the number of bound antibody molecules per cell line (estimating receptor count). Differential MFI (Mean Fluorescence Intensity) and ABC values are shown in the table and visualized in the overlay histogram for the four stable cell lines tested enabling the ranking of high → low target expression and therefore selection of cell lines.

Figure 2. CHO cells stably expressing human or cyno targets of interest were tested in the receptor quantification assay using the Quantum Simply Cellular® kit (Bangs Laboratories). Flow cytometry analysis was performed on the Attune™ NxT flow cytometer.

The receptor quantification beads measure antibody binding capacity (ABC) and were stained alongside cells with a positive control antibody; a standard curve is generated, enabling interpolation of fluorescence intensity to the number of bound antibody molecules per cell line (to estimate receptor count).

Differential MFI (mean fluorescence intensity) and ABC values are shown in the table and visualized in the overlay histogram for the four stable cell lines tested, enabling the ranking of high → low target expression and therefore selection of cell lines. Image Credit: Image courtesy of Zaynab Isseljee et al., in partnership with ELRIG (UK) Ltd.

3. HT flow cytometry sera screening

During the hit identification stage, mice receive immunizations over several weeks with immunogens that present the target of interest. Bleeds are collected, and sera screened by ELISA and flow cytometry across this period to identify mice showing an immune response against the target.

  • Validated target-expressing cell lines are used in an HT flow cytometry sera screening assay to assess sera collected from the bleeds for target binding ability (Figure 3).
  • Collating and ranking the flow and ELISA sera screening data allows selection of mice showing the highest antibody titers for progression to single B-cell sorting, sequencing, and analysis stage

HT Flow Cytometry Sera Screening Assay performed on the Attune™ NxT flow cytometer, with binding of sera to target cells detected using a secondary PE labelled antibody. (A) Representative overlay histograms illustrating the binding of sera dilutions to human and cyno target cells. No sera binding seen against empty vector cells. (B) Final bleeds show the differential binding of sera dilutions to human and cyno target cells. Antibody titers for recombinant target protein and target overexpressing cells also showed similar ranking

HT Flow Cytometry Sera Screening Assay performed on the Attune™ NxT flow cytometer, with binding of sera to target cells detected using a secondary PE labelled antibody. (A) Representative overlay histograms illustrating the binding of sera dilutions to human and cyno target cells. No sera binding seen against empty vector cells. (B) Final bleeds show the differential binding of sera dilutions to human and cyno target cells. Antibody titers for recombinant target protein and target overexpressing cells also showed similar ranking

Figure 3. HT flow cytometry sera screening assay performed on the Attune™ NxT flow cytometer, with binding of sera to target cells detected using a secondary PE-labeled antibody. (A) Representative overlay histograms illustrating the binding of sera dilutions to human and cyno target cells. No sera binding was seen against empty vector cells.

(B) Final bleeds show the differential binding of sera dilutions to human and cyno target cells. Antibody titers for the recombinant target protein and for target-overexpressing cells also showed similar rankings. Image Credit: Image courtesy of Zaynab Isseljee et al., in partnership with ELRIG (UK) Ltd.

4a. HT flow cytometry antibody screening

A multiplex flow cell binding assay was developed and used at the HT screening and lead panel characterization stages.

  • The selected cell lines produced in reagent generation are multiplexed and used to screen a large panel of antibodies to identify positive hits.
  • Recent workflow enhancements have further strengthened LifeArc's flow-screening capabilities through the integration of a mobile robotic automation platform (FlexPod®, from HighRes®).

Multiplex flow cytometry assay set-up

Using flow cytometry to support antibody discovery workflows

Image Credit: Image courtesy of Zaynab Isseljee et al., in partnership with ELRIG (UK) Ltd.

4b. HT flow antibody screening analysis

Following screening of an HT antibody panel in the multiplex flow cell binding assay, data analysis includes a gating strategy, verification of control binding, and generation of heat maps indicating positive binders for each species (Figure 4).

(A) Representative gating strategy illustrated with dot plots. Detected bound antibody is observed against each individual cell population. (B) Control antibodies show expected cell binding against human and cyno target cell lines, shown as MFI fold over background (empty vector cells). (C) Representative heat maps of the MFI values highlight positive binders (purple shaded wells) within assay plates for each cell line tested.

Figure 4. (A) Representative gating strategy illustrated with dot plots. The detected bound antibody is observed for each individual cell population. (B) Control antibodies show expected cell binding against human and cyno target cell lines, shown as MFI fold over background (empty vector cells). (C) Representative heat maps of MFI values highlight positive binders (purple-shaded wells) across assay plates for each cell line tested. Image Credit: Image courtesy of Zaynab Isseljee et al., in partnership with ELRIG (UK) Ltd.

4c. HT flow antibody screening data visualization

MFI values shown in heat maps are exported to generate scatter plots to visualize cross-reactive and non-cross-reactive binders, as well as correlations between screening runs.

Collation of flow cytometry data with other screening datasets is then performed to enable down-selection from the HT panel to mid-scale and lead candidates.

Multiplex flow cell binding data visualization demonstrated good correlation of MFI fold between

Figure 5. Multiplex flow cell binding data visualization demonstrated good correlation of MFI fold between n=1 and n=2 screening runs for (A) human target cells and (B) cyno target cells. 66% and 61% mAbs showed cross-reactive binding to both human and cyno target cells in n=1 and n=2, respectively. Flow-binding data also correlated well with SPR kinetic data. Image Credit: Image courtesy of Zaynab Isseljee et al., in partnership with ELRIG (UK) Ltd.

Summary

  • Flow cytometry assays shown here are key for validating and characterizing cell line target expression and receptor quantification, enabling high-throughput serum and antibody screening. These flow cytometry workflows enable the discovery of antibody therapeutics using LifeArc's B-SMART™ antibody discovery platform.
  • Integration of the FlexPod automation has increased flow cytometry screening efficiency threefold, increasing the screening from 100 to 300 antibodies, improving consistency and supporting the generation of high-throughput data to enhance data-driven decisions, whilst maintaining assay quality and saving significant time.
  • LifeArc's Flow Specialist Group continues to seek future opportunities to expand flow cytometry applications and capabilities within LifeArc's discovery pipeline

Reference

Multiplex cell binding protocol adapted from Intellicyt® App Note Multiplexed Antibody Characterization.

About LifeArc

LifeArc® is a medical research charity making life science life changing, transforming promising life science ideas into medical breakthroughs that change patients’ lives. We are self-funding and specialize in early-stage translation – advancing lab-based scientific discoveries to a point at which they can be developed into the next generation of diagnostics, treatments and cures. We have been doing this for more than 25 years and our work has resulted in a diagnostic for antibiotic resistance and four licensed medicines. This includes Keytruda®(cancer), Actemra® (rheumatoid arthritis), Tysabr® (multiple sclerosis) and Entyvio® (Crohn’s disease) and a test for antimicrobial resistance.

About ELRIG (UK) Ltd.

The European Laboratory Research & Innovation Group (ELRIG) is a leading European not-for-profit organization that exists to provide outstanding scientific content to the life science community. The foundation of the organization is based on the use and application of automation, robotics, and instrumentation in life science laboratories, but over time, we have evolved to respond to the needs of biopharma by developing scientific programs that focus on cutting-edge research areas that have the potential to revolutionize drug discovery.

Comprised of a global community of over 12,000 life science professionals, participating in our events, whether it be at one of our scientific conferences or one of our networking meetings, will enable any of our community to exchange information, within disciplines and across academic and biopharmaceutical organizations, on an open access basis, as all our events are free of charge to attend!

Our values

Our values are to always ensure the highest quality of content, that content will be made readily accessible to all, and that we will always be an inclusive organization, serving a diverse scientific network. In addition, ELRIG will always be a volunteer-led organization, run by and for the life sciences community, on a not-for-profit basis.

Our purpose

ELRIG is a company whose purpose is to bring the life science and drug discovery communities together to learn, share, connect, innovate, and collaborate on an open-access basis. We achieve this through the provision of world-class conferences, networking events, webinars, and digital content.


Sponsored Content Policy: News-Medical.net publishes articles and related content that may be derived from sources where we have existing commercial relationships, provided such content adds value to the core editorial ethos of News-Medical.net, which is to educate and inform site visitors interested in medical research, science, medical devices, and treatments.

Last Updated: Nov 2, 2026

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    LifeArc. (2026, November 02). Using flow cytometry to support antibody discovery workflows. News-Medical. Retrieved on November 07, 2026 from https://www.news-medical.net/health/Using-flow-cytometry-to-support-antibody-discovery-workflows.aspx.

  • MLA

    LifeArc. "Using flow cytometry to support antibody discovery workflows". News-Medical. 07 November 2026. <https://www.news-medical.net/health/Using-flow-cytometry-to-support-antibody-discovery-workflows.aspx>.

  • Chicago

    LifeArc. "Using flow cytometry to support antibody discovery workflows". News-Medical. https://www.news-medical.net/health/Using-flow-cytometry-to-support-antibody-discovery-workflows.aspx. (accessed November 07, 2026).

  • Harvard

    LifeArc. 2026. Using flow cytometry to support antibody discovery workflows. News-Medical, viewed 07 November 2026, https://www.news-medical.net/health/Using-flow-cytometry-to-support-antibody-discovery-workflows.aspx.

Comments

The opinions expressed here are the views of the writer and do not necessarily reflect the views and opinions of News Medical.
Post a new comment
Post

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.