REPRISE II trial: Lotus Valve System meets primary endpoint in severe aortic valve disease patients

NewsGuard 100/100 Score

Boston Scientific Corporation (NYSE: BSX) reports positive results from a pre-specified analysis of the first 60 patients enrolled in the REPRISE II trial evaluating the performance and safety of the Lotus™ Valve System in symptomatic patients with severe aortic valve disease considered at high risk for surgical valve replacement. The Lotus Valve System is the first transcatheter aortic valve replacement (TAVR) device of its kind with an Adaptive Seal™ that is designed to minimize aortic regurgitation (leaking), a proven predictor of mortality, and is both fully repositionable and retrievable prior to release. Data were presented today as a late-breaking clinical trial at the EuroPCR Scientific Program in Paris by Professor Ian Meredith , director of MonashHeart, at Monash Medical Centre in Melbourne, Australia, and principal investigator of the REPRISE II trial. The data demonstrated that the Lotus Valve System met the primary performance endpoint for the first 60-patient cohort and was implanted successfully in all patients (60/60 patients) with no case of severe paravalvular regurgitation.

REPRISE II is an ongoing prospective, single-arm study that has completed enrollment of 120 patients at fourteen sites in Australia, France, Germany and the United Kingdom. All patients had severe symptomatic aortic stenosis and were considered at high risk for surgical valve replacement. REPRISE II is being extended to enroll an additional 130 patients at twenty sites in Australia and Europe.

Results

  • The primary device performance endpoint was met as the 30-day mean aortic valve pressure gradient of 11.28+5.23 mmHg with a one-sided upper confidence bound of 13.09 was significantly less (P<0.0001) than the performance goal of 18 mmHg.
  • The primary safety endpoint, defined as all-cause mortality at 30 days, was 1.7 percent.
  • Independent core lab assessment of paravalvular aortic regurgitation at 30 days indicated no severe regurgitation and one case of moderate regurgitation (1.9 percent).  In 79.2 percent of patients there was trace or no paravalvular regurgitation at 30 days.
  • No instances of non-study valve implantation, unplanned use of cardiopulmonary bypass, valve embolization, valve-in-valve or ectopic valve placement occurred.

"The ability to initially position the Lotus valve very precisely and, if needed, to easily reposition or indeed fully retrieve the valve provide the operator with remarkable control," said Professor Meredith. "When combined with immediate and near complete elimination of aortic regurgitation, even in patients who began with moderate or severe aortic regurgitation, the results are impressive and reflect very favorably on the unique features of the Lotus technology."

One-year results from REPRISE I, a prospective, single-arm feasibility study on patients with severe symptomatic aortic stenosis conducted in Australia, were also presented today by Professor Meredith. Data demonstrated sustained safety and performance of the Lotus valve out to one year, with no new major adjudicated events as defined by the Valve Academic Research Consortium (VARC), and no moderate or severe paravalvular aortic regurgitation in any patients.

"The new results from our Lotus trial program, and particularly from REPRISE II, underscore the promising unique technology behind the Lotus Valve System," said Keith Dawkins, M.D., global chief medical officer, Boston Scientific. "These features help to simplify the implantation procedure and may lead to improved clinical outcomes, showing how the Lotus valve can offer a new treatment alternative for patients with severe aortic valve disease considered being at high risk for surgical valve replacement."

The Lotus Valve System is an investigational device worldwide, limited by applicable law to investigational use and not available for sale. The device was developed by Sadra Medical, which Boston Scientific acquired in 2011. 

The results of the REPRISE trials are expected to be used to support CE mark and other international regulatory approvals.

Comments

The opinions expressed here are the views of the writer and do not necessarily reflect the views and opinions of News Medical.
Post a new comment
Post

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.

You might also like...
Higher consumption of ultra-processed foods linked to elevated mortality risk