Defibtech initiates voluntary recall of DDU-100 series semi-automatic external defibrillators

Defibtech, LLC is initiating a worldwide voluntary recall of certain DDU-100 series semi-automatic external defibrillators (AEDs) sold under the Lifeline AED and ReviveR AED brand names, including 65,885 AEDs distributed in the United States. This recall affects only DDU-100 Series AEDs shipped with 2.004 software or earlier. This corrective action addresses two possible conditions, which in rare cases may cause an affected AED to cancel shock during the charging process and not provide therapy which may result in failure to resuscitate the patient. Both conditions are not detectable by the periodic self test.

Condition 1: In rare instances, the AED may cancel charge in preparation for a shock. Based on field data, the odds of an affected AED having this happen are less than a 1 in 400,000 chance per month for any given AED.

A subset of AEDs (less than 11%) that are affected by Condition 1 are also affected by Condition 2: In rare instances, the AED may cancel charge in preparation for a shock in very high humidity conditions. The only reported cases were in environments of greater than 95% relative humidity or condensing conditions. Based on field data, the odds of an affected AED having this happen are less than a 1 in 250,000 chance per month for any given affected AED.

Defibtech will provide customers with a free software upgrade to address these issues. The correction to the AED will be able to be performed at the location where the AED is deployed.

Because both of these conditions occur very rarely, it is recommended that customers keep their AEDs in service until they have performed the software upgrade. Full instructions and recommendations are being mailed to affected customers. This customer notification, as well as instructions on determining whether an AED is affected, can also be found on the www.defibtech.com/fa11 web page. For additional information regarding this recall, please refer to the above referenced web page, contact your distributor, or contact Defibtech at [email protected], 1-877-453-4507 or 1-203-453-4507.

The AEDs affected by this recall have been distributed globally to fire departments, EMS, health clubs, schools, and other organizations. The Food and Drug Administration (FDA) has determined that this action is a Class I recall. Any adverse reactions experienced with the use of this product and/or quality problems should also be reported to the FDA's MedWatch Program by phone at 1-800- FDA-1088, or on the MedWatch website at www.fda.gov/medwatch.

Comments

The opinions expressed here are the views of the writer and do not necessarily reflect the views and opinions of News Medical.
Post a new comment
Post

Sign in to keep reading

We're committed to providing free access to quality science. By registering and providing insight into your preferences you're joining a community of over 1m science interested individuals and help us to provide you with insightful content whilst keeping our service free.

or

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.

You might also like...
New meta-analysis challenges claims of athletic advantage for transgender women